Air Monitoring Alarm Response
Air Monitoring Instrument
Thermo Scientific Model 5030 Sharp Monitor
Date and time of alarm
12/1/2024 1:00 AM
to
12/31/2024 12:59 PM
Weekend
Non-working hours
Active contracting onsite
Alarm was caused by
Loss of power
Loss of radio communication
Loss of cellular connection
Out of calibration
Air quality, traffic, site maintenance, insects
Rain
Air unit malfunction
Other:
Off line
Name of technician who responded to the alarm
Scott Peterson
Was an email sent to the site technician?
Yes
No
Were dredging or construction occurring during the alarm?
Yes
No
Were corrective actions taken?
Yes
No
Description of action taken
During the month of December, error air-units offline. Found antenna mounted to fence pole hose clamps were not tight. The wind was pushing the antenna out of alignment with base antenna. Tightened the hose clamps but may need to replace with heavy duty style clamp.